Axial source and focusing
- EI source
- Einzel lens
- Quadrupole
An Einzel lens couples the source to the analyzer along a shared axis. A direct source mounting option is also available when the selected layout calls for a shorter stack.
Products / Systems / Flange-mounted mass filter probes
Ardara system families
Integrate ion generation and quadrupole mass selection into your vacuum instrument. Match the source position, beam path, analyzer and mounting flange to the experiment.

Ionization inside your instrument
An electron-impact source can ionize a directed neutral molecular beam or residual gas already inside the vacuum chamber. The source and ion optics then couple the generated ions into the quadrupole.
Both the Gen 7 / Dumbo and Jumbo ionizers can mount directly to the quadrupole housing, to an Einzel lens, or to a deflector.
An Einzel lens couples the source to the analyzer along a shared axis. A direct source mounting option is also available when the selected layout calls for a shorter stack.
A quadrupole deflector turns the ion path toward the analyzer. Cross-beam layouts allow the neutral beam and analyzer axes to be arranged separately, useful when the incoming material should not pass through the analyzer and detector.
Conceptual ion paths. Source mounting, neutral-beam direction and final port routing are established in the configuration drawing. Deflector-to-deflector arrangements also support coaxial ionization and transmission.
The newest slim Gen 7 / Dumbo is the smaller standard source across the three quadrupole sizes. Jumbo provides a larger ionization region and is the preferred pairing for the 12 and 20 mm quadrupoles.
Select the source volume, mounting arrangement and ion optics together for the beam and chamber.
The documented molecular-beam source family uses removable stainless-steel ion-region and optical components for maintenance and cleaning. Leave space to reach the source and its connections when planning the chamber.
Explore ionizers and opticsMechanical options
Match rod size, rod length and ion optics to the available space and the intended experiment.
| Quadrupole size | Rod length | Prefilters | Mounting flange |
|---|---|---|---|
| 9.25 mm | 8 inches | Optional | 6-inch or 8-inch CF |
| 12 mm | 8 or 12 inches | Optional | 6-inch or 8-inch CF |
| 20 mm | 8 inches | Optional | 6-inch or 8-inch CF |
Overall quadrupole length is approximately 0.100 inch greater than the rod length. The ion source, lenses, deflectors and flange hardware add to the complete system envelope.
Explore the assembly
The representative assembly combines a Jumbo ionizer, Einzel lens stack and 12 mm quadrupole on an 8-inch CF mounting assembly. Explore four fixed views of this configuration.

Isometric view · 100%
The ionizer forms ions from the incoming molecular beam or residual gas.
The axial lenses couple the source to the analyzer entrance.
The quadrupole provides mass selection in the illustrated reference build.
Mechanical integration
Use the front and side reference views to discuss the arrangement, then request a dimensioned elevation matched to your configuration.
CF8 / 12 mm Jumbo ionizer and Einzel lens reference. CAD reference views; request a dimensioned elevation for the selected configuration.
Confirm the flange interface, insertion space and complete source-to-flange envelope before finalizing the chamber. Allow access for feedthroughs, connections and source servicing.
Request a dimensioned elevationThe CF6 / 9.25 mm example combines a smaller ionizer with a deflector.
Complete the instrument
Plan the analyzer, ion optics, detector and source controls as a matched configuration.
Match the quadrupole and RF supply to the intended mass-selection conditions.
View component details →Plan filament supply and source operation together with the chosen ionizer.
View component details →Choose detector hardware and bias control around the measurement system.
View component details →Reference documentation
Use these existing Ardara documents alongside a configuration-specific drawing. Confirm the document and assembly revisions for the proposed build.
Reference ionizer layouts, ion-region sizes and coupling options.
Open PDF → (PDF; opens in a new tab)Reference for the controller, RF driver and inductance module. Confirm the configuration matched to your quadrupole.
Open PDF → (PDF; opens in a new tab)PDF documents open in a new tab. Browse all manuals · Request documents
Yes. Prefilters are available on 20 mm quadrupoles, as well as the 9.25 and 12 mm sizes. Confirm the required configuration with Ardara.
It describes the rod length, and is offered for the 12 mm quadrupole. Standard rods are 8 inches. Source and mounting hardware determine the complete probe envelope.
The newest slim design is the current Gen 7 / Dumbo ionizer. It supports all three quadrupole sizes. Jumbo is the preferred pairing with the 12 and 20 mm quadrupoles.
Yes. The source and deflector arrangements include perpendicular ionization, as well as coaxial ionization and transmission. Establish the final port and optics layout with a configuration drawing.
Provide the chamber layout, incoming beam or gas, expected operating environment, accessible mounting ports and detector requirements. These inputs help select the smaller Gen 7 / Dumbo or Jumbo source and its optics.
The probe family combines a mounting flange, analyzer and selected source or optics for insertion into a chamber. The in situ beamline family uses a radial-flange architecture to mount the mass filter within the beamline.
These probe arrangements can generate ions from a molecular beam or residual gas inside vacuum. An atmospheric-pressure inlet adds the source interface and staged vacuum transport needed to bring externally formed ions into the instrument.
Assembly reference gallery
Explore example layouts and discuss the required configuration with Ardara.
10 reference assemblies
Configuration example
A CF6-mounted 9.25 mm quadrupole configuration with smaller molecular-beam ionizer with quadrupole deflector.

Configuration example
A CF6-mounted 9.25 mm quadrupole configuration with smaller molecular-beam ionizer with Einzel lens stack.

Configuration example
A CF8-mounted 12 mm quadrupole configuration with quadrupole entrance lens only.

Configuration example
A CF8-mounted 12 mm quadrupole configuration with Jumbo ionizer with quadrupole deflector.

Configuration example
A CF8-mounted 12 mm quadrupole configuration with Jumbo molecular-beam ionizer with Einzel lens stack.

Configuration example
A CF8-mounted 12 mm quadrupole configuration with smaller molecular-beam ionizer with Einzel lens stack.

Configuration example
A CF8-mounted 12 mm quadrupole configuration with quadrupole entrance lens only.

Configuration example
A CF8-mounted 20 mm quadrupole configuration with Jumbo molecular-beam ionizer with Einzel lens stack.

Configuration example
A CF8-mounted 20 mm quadrupole configuration with Jumbo ionizer with quadrupole deflector.

Configuration example
A CF8-mounted 9.25 mm quadrupole configuration with Jumbo ionizer with quadrupole deflector.