Bring atmospheric-pressure ions into a vacuum instrument with a matched source interface, ion transport and differential pumping. These reference builds use electrospray ionization (ESI).
Integrated ESI / CF8 / 12 mm quadrupole reference assembly. The source interface and analyzer are separate functional stages.
Follow the ion path
From electrospray to a mass-selected signal
The documented 12 mm ESI system combines a sprayer and heated transfer capillary with an ion funnel, a differentially pumped ion guide, a quadrupole and a detector. Each stage has a distinct role.
1
Electrospray sprayer
Forms ions at the atmospheric-pressure source.
2
Heated transfer capillary
Provides the entrance from the source into vacuum.
3
Ion funnel
Collects the entering beam and passes it to downstream optics.
4
Differential ion guide
Transports ions through the pumped vacuum stages.
5
Quadrupole mass filter
Provides mass selection in the integrated reference build.
6
Detector
Receives the selected ion beam for signal collection.
Functional sequence of the documented reference system. The sketch describes the ion path; it does not define the physical scale or chamber layout.
Vacuum and transport work together. Pumping connections, conductance between stages and ion optics are planned as one interface. Confirm the pumps, gauges and control electronics included in the proposed build.
Define the scope
Inlet assembly or integrated system
The ESI base reference illustrates the source interface. The integrated ESI / Q12 reference also illustrates mass selection and detector-side hardware; confirm the scope of the proposed build.
Inlet assembly
Connect to an existing analyzer
Start with the atmospheric source interface and ion-transport stages. Match the downstream vacuum connection, beam path and controls to the instrument that receives the ions.
Source position and sample delivery
Transfer capillary, heating and ion transport
Pumping interfaces and service access
Integrated reference
Include mass selection and detection
The illustrated ESI / Q12 build adds the quadrupole and detector-side hardware. Source control, transport supplies, analyzer drive and data acquisition are planned as a complete chain.
12 mm quadrupole reference architecture
Matched detector and control electronics
Instrument-level mechanical integration
Explore the assembly
Explore the integrated ESI / quadrupole layout
The existing R2 reference combines the atmospheric-pressure source interface with an 8-inch CF / 12 mm quadrupole assembly.
Isometric view · 100%
Mechanical integration
Plan the mounting envelope
Use the front and side reference views to discuss the arrangement, then request a dimensioned elevation matched to your configuration.
Front reference Open full-size →
Side reference Open full-size →
Integrated ESI / CF8 / 12 mm reference. CAD reference views; request a dimensioned elevation for the selected configuration.
Bring the chamber and probe together
Confirm the flange interface, insertion space and complete source-to-flange envelope before finalizing the chamber. Allow access for feedthroughs, connections and source servicing.
Is the ESI base assembly a complete mass spectrometer?
No. It is a source-interface assembly. The separate integrated example includes the quadrupole and detector-side hardware.
Can an inlet be integrated with an existing instrument?
Discuss the downstream vacuum interface, pumping, ion optics and control requirements with Ardara so that the inlet is matched to the instrument.
What should accompany an inlet configuration request?
Include the source and sample delivery, expected gas or solvent load, downstream vacuum interface, pumping arrangement and available controls. A chamber sketch helps establish the mounting and service envelope.