Overview
The PSRF-128 adjusts its resonant frequency to the connected capacitive load. Manual and external RF-amplitude control, a front-panel RF display and vacuum interlock support integration with RF-only ion-transport structures.
Product details
| Operating role | RF-only ion guides, including quadrupoles and multipoles |
|---|---|
| Frequency generation | Self-oscillating; frequency depends on the connected load and configuration |
| External amplitude command/readback | DB9 interface |
| Controller-to-RF-head connection | DB25 interface |
| RF display | Peak-to-peak or zero-to-peak scale according to the unit label |
| Control | Manual/external selection; vacuum interlock input |
Details describe the documented configuration. Confirm the supplied hardware and options with Ardara.
Integration
Provide total capacitive load including cables, desired RF amplitude, electrode geometry and DC bias requirements. Frequency, voltage limit and display scaling must follow the supplied unit’s configuration markings.
Documentation
Contact Ardara for the manual and configuration documentation that applies to your Model 128 unit. Include its serial number when requesting support.
Discuss Model 128