Overview
The QPS-129 provides mass-command, pole-bias and resolution controls for a matched quadrupole RF system. Its crystal-fixed-frequency architecture is tuned to the load through the associated RF hardware.

Product details
| Operating modes | RF-only and resolving RF/DC |
|---|---|
| Front-panel controls | Mass command, pole bias, low-mass resolution and high-mass resolution |
| Pole bias | −50 to +50 V DC in the documented configuration |
| Frequency architecture | Crystal-fixed frequency; resonant tuning matched to the load |
| RF-head interface | DB25 interconnect |
| Control | Manual/external selection; vacuum interlock |
Details describe the documented configuration. Confirm the supplied hardware and options with Ardara.
Integration
Select the controller and RF head as a matched configuration. Confirm the analyzer geometry, load, operating frequency, calibration and supplied cable set. Model 129 and Model 129A are distinct configuration references; do not assume interchangeability.
Documentation
Contact Ardara for the manual and configuration documentation that applies to your Model 129 unit. Include its serial number when requesting support.